360 V Bidirectional DC Power Supply

360 V Bidirectional DC Power Supply

360 V bidirectional, regenerative, programmable DC power supply. High-current 360 V class for low-voltage, high-amperage pack and module test. Source and sink in one unit, with energy regenerated to the grid at up to 93%.

Bidirectional93% regen≤5 ms0.02% F.S.CC/CV/CP/CRSCPI · Modbus · CANopen

Models in this class

Every model below is a bidirectional source + regenerative load. Request a quote on any configuration — or ask us to size a parallel system.

VoltageCurrentPowerFormModel
360 V±80 A14 kW3UN35514-360-80Quote →
360 V±160 A22 kW3UN35522-360-160Quote →
360 V±240 A32 kW3UN35532-360-240Quote →
360 V±240 A42 kW3UN35542-360-240Quote →
360 V±480 A84 kW6UN35584-360-480Quote →
360 V±720 A126 kW22UN355126-360-720Quote →
360 V±960 A168 kW22UN355168-360-960Quote →
360 V±1200 A210 kW32UN355210-360-1200Quote →
360 V±1440 A252 kW32UN355252-360-1440Quote →
360 V±1680 A294 kW32UN355294-360-1680Quote →
360 V±1920 A336 kW42UN355336-360-1920Quote →
360 V±2160 A378 kW42UN355378-360-2160Quote →
360 V±2400 A420 kW42UN355420-360-2400Quote →

View & quote this product page →

Capabilities

Bidirectional source + regenerative load, plus optional battery simulation (NS81000, 7 model libraries + custom curves) and PV array simulation (NS91000, EN50530 / Sandia static & dynamic MPPT).

Technical specifications

Technical specifications (representative)
Output voltage range0 – rated (360 / 500 / 750 / 1000 / 1500 / 2250 V)
Output currentBidirectional ± rated (to 240 A / unit, to 2400 A in cabinet)
Output powerBidirectional ± rated (14 kW–420 kW; MW via parallel)
Operating quadrantsDual-quadrant (source + regenerative sink), seamless crossover
Regenerative efficiencyUp to 93% (energy returned to grid)
CV setting accuracy≤ 0.02% + 0.02% F.S.
CC / CP setting accuracy≤ 0.1% + 0.1% F.S.
Voltage readback accuracy≤ 0.02% + 0.02% F.S.
Dynamic responseVoltage rise/fall ≤ 5 ms (representative 3U); transient recovery < 4.5 ms (10→60%)
Load modesCC / CV / CP / CR, with CC&CV priority
SimulationBattery (NS81000, 7 libraries + custom) · PV I-V/P-V (NS91000, EN50530/Sandia)
ProtectionsOVP / OCP / OPP / UVP / UCP
Power factor0.99
AC inputThree-phase 340–480 VAC, 47–63 Hz
InterfaceLAN / RS232 / RS485 / CAN — SCPI, Modbus-RTU, CANopen
Display6.8″ LCD
Operating environment0–50 °C (derate > 35 °C); storage −10–70 °C; < 2000 m; 5–90% RH non-condensing
Standards / qualityCE · IEC 61010-1 · ISO 17025-traceable calibration (on request)

Need this configured for your test?